TECHNICAL MANUAL  
MODEL: 61-362  
The Service Information provides the following information  
• Precautions and safety information  
• Specifications  
• Performance test procedure  
• Calibration and calibration adjustment procedure  
• Basic maintenance (replacing the battery and fuses)  
Form number: 61362  
Rev 1. date: May 2001  
Form number 61362  
rev 1 May 2002  
 
Page 1  
Introduction  
Warning  
To avoid shock or injury, do not perform the verification tests or calibration  
procedures described in this manual unless you are qualified to do so.  
The information provided in this document is for the use of qualified personnel  
only.  
Caution  
The 61-362 serials contain parts that can be damaged by static discharge.  
Follow the standard practices for handling static sensitive devices.  
For additional information about IDEAL INDUSTRIES and its products,  
and services, visit IDEAL INDUSTRIES web site at:  
SAFETY  
Review the following safety precautions to avoid injury and prevent damage to this product or any  
products connected to it. To avoid potential hazards, use the product only as specified.  
It is recommended that you read through the Operation or User manual before starting. Not all Caution,  
Warning, or Danger precautions are listed in this manual.  
CAUTION.  
These statements identify conditions or practices that could result in damage to the equipment or other  
property.  
WARNING.  
These statements identify conditions or practices that could result in personal injury or loss of life.  
Specific precautions  
Use proper Fuse. To avoid fire hazard, use only the fuse type and rating specified for this product.  
Do not operate without covers. To avoid personal injury, do not apply any voltage or current to the  
product without the covers in place.  
Electric overload. Never apply a voltage to a connector on the product that is outside the range specified  
for that connector.  
Avoid electric shock. To avoid injury or loss of life, do not connect or disconnect probes or test leads  
while they are connected to a voltage source.  
Do not operate in wet/damp conditions. To avoid electric shock, do not operate this product in wet or  
damp conditions.  
Form number 61362  
rev 1 May 2002  
 
Page 2  
General specifications  
Model 61-362  
Characteristics  
Display  
Description  
3 ½ Digit LCD display  
3200 count, maximum reading 3199  
32 segment analog bar graph  
“OL” displayed  
Display Count  
Over range Indication  
Sampling Rate  
2 time/second  
Operating Environment:  
Relative Humidity  
Temperature Coefficient  
Storage Environment:  
Power source:  
0°C to 50°C (32°F to 122°F) 70%RH  
0.05X x (accuracy) per °F (32°F to 65°F, 85°F to 122°F  
-20°C to 60°C (0°F to 140°F) at <70 relative humidity  
9V Battery (NEDA 1604)  
Battery Live:  
200 hours typical (alkaline)  
Low Battery Indicator:  
Auto Power Off mode  
A protection Fuse  
symbol indicates low battery voltage  
Approximately 25 minutes  
0.5A/250V fast acting fuse Type LA-3895  
10A/600V fast acting fuse, Type LA-3897  
Dimensions  
Weight:  
Safety  
5.75” H X 2.75” W X 1.5” D {without holster}  
Approximately 18.0 oz. including battery  
UL1244, and Design to comply with IEC 1010-1 Cat III  
RANGES and ACCURACY SPECIFICATION  
61-362  
Function Setting  
Ranges  
Accuracy  
3.200V/32.00V/320.0V/ 600V  
50Hz to 500Hz  
2.0% ± 4 digits  
AC Voltage  
320.0mV/32.00V/320.0V/600V  
3.200V  
320µA/3200 µA/32.00mA/320mA  
10A,  
50Hz to 500Hz  
1.2% ± 1digit  
0.8% ± 1digit  
2.5% ± 4 digits  
3.5% ± 4 digits  
DC Voltage  
AC Current  
DC Current  
320µA/3.20mA/32.00mA  
2.0% ± 1 digit  
10A,  
320.0Ω  
3.200k/32.00k/320.0k  
3.200MΩ  
32.00MΩ  
beep on < 100Ω  
2V DC max  
3.0% ± 1 digit  
2.0% ± 3 digits  
1.5% ± 3 digits  
2.5% ± 3 digits  
5.0% ± 5 digits  
Not Specified  
10% ± 2 digit  
Resistance  
Continuity  
Diode Check  
AC Converter:  
Overload Protection:  
Average responding, RMS Calibrated to Sine Wave  
AC and DC Volts: 600V DC or AC rms .  
Resistance, Diode, Continuity: 500V DC or AC rms  
mA input: .5A/250V DC/AC rms  
Amps input: 10A/600V DC/AC rms  
Form number 61362  
rev 1 May 2002  
 
Page 3  
PERFORMANCE VERIFICAITONS  
Perform the following analysis, if the meter conforms to the limits listed in Table 1 the meter is functioning  
correctly. If the meter does not conform to any of the listed limits the calibration procedure must be  
performed.  
Performance Verification Preparation  
1. Turn on the Calibrator, allow calibrator to warm up. Temperature Stabilization  
should be reached after 30 minutes.  
2. Remove battery cover and using a calibrated meter to ensure the battery measures a minimum of 7.5 V  
DC. If the battery measures under 7.5V DC, replace the battery before beginning the performance test.  
3. Input the values listed in Table 1 for the Standard 61-362  
Table 2 Performance Verification 61-362  
Function Setting /Range  
ACV 3.2V  
ACV 3.2V  
ACV 32  
ACV 32  
ACV 320  
ACV 320  
ACV 600  
ACV 600  
DCV 320mV  
DCV 3.2V  
DCV 32V  
DCV 320  
DCV 600  
Input  
Low Limit  
2.936  
2.936  
29.36  
29.36  
293.6  
293.6  
486  
High Limit  
3.064  
3.064  
30.64  
30.64  
306.4  
306.4  
514  
3.000V @ 50Hz  
3.000V @ 500Hz  
30.00 @ 50Hz  
30.00 @ 500Hz  
300V AC @ 50Hz  
300V AC @ 500Hz  
500V AC @ 50Hz  
500V AC@ 60Hz  
300mV DC  
3.000V DC  
30.00V DC  
300.0V DC  
500V DC  
486  
514  
296.3  
2.975  
29.63  
296.3  
493  
303.7  
3.025  
30.37  
303.7  
507  
A DC 320µA  
A DC 3200 µA  
A DC 32mA  
A DC 320mA  
A DC 10  
300µA  
3000µA  
30mA  
300mA  
293.9  
2939  
29.39  
293.9  
8.70  
306.1  
3061  
30.61  
306.1  
9.30  
9.00 DCA  
A AC 320µA  
A AC 320µA  
A AC 3200µA  
A AC 320oµA  
A AC 32mA  
A AC 32mA  
A AC 320mA  
A AC 320mA  
A AC 10A  
A AC 10A  
320  
2k  
32k  
320k  
2M  
32M  
Diode Test  
Continuity Test  
Continuity Test  
300µA @ 50Hz  
300µA @ 500Hz  
3000µA @ 50Hz  
3000µA @ 500Hz  
30mA @ 50Hz  
30mA @ 500Hz  
300mA @ 50Hz  
300mA @500Hz  
9A @ 50Hz  
9A @ 500Hz  
300.0  
3.000k  
30.00k  
300.0k  
3.000M  
30.00M  
500mV  
120Ω  
80Ω  
292.1  
292.1  
2921  
307.9  
307.9  
3079  
3079  
30.79  
30.79  
307.9  
307.9  
9.36  
2921  
29.21  
29.21  
292.1  
292.1  
8.64  
8.64  
9.36  
297.7  
2.950  
29.52  
295.2  
2.922  
28.45  
.448  
306.3  
3.048  
30.48  
304.8  
3.078  
31.55  
.552  
Beep off  
Beep on  
Form number 61362  
rev 1 May 2002  
 
Page 4  
CALIBRATION  
Calibration Preparation  
Required Equipment  
The class of calibrator or equipment should have an accuracy that exceeds, by an expectable ratio the  
accuracy of this instrument.  
1. Turn on the Calibrator, allow calibrator to warm up. Temperature Stabilization  
should be reached after 30 minutes.  
2. Disconnect the test leads and turn the range switch to “OFF”.  
3. Remove the screw holding the bottom case cover, just above the battery cover.  
4
The case bottom is secured to the case top by two internal snaps. (at the LCD end).  
lift up on the battery end until the case un-snaps.  
5
Using a calibrated meter ensure the battery measures a minimum of 7.5 V dc.  
If the battery measures under 7.5V DC, replace the battery.  
Calibration Procedure  
It is recommended that all IDEAL meters undergo the following calibration procedure on  
an annual basis.  
61-362 Calibration Procedure.  
Calibration  
1. Set the Function/Range Switch to the “3.200V DC” position.  
2. Set the output of the DC calibrator for 3.000V and connect it to the “V-Ohm” and “COM”  
input terminals.  
3. Adjust VR1 (VR 200 ohm) until the display reads 3.000V +/- 1 digit.  
Note : This is the only adjustments required for the 61-362. Calibration is complete.  
Battery Replacement (refer to Figure 2)  
1. Disconnect the test leads from any circuit under test and turn off meter.  
2.Remove the three screws for the back case cover.  
3. Remove battery from compartment noting the “+” and “- “ position of the Battery terminals.  
4. Remove the Battery plastic sleeve and place it on the new battery  
(Damage can occur to circuit if Plastic sleeve is not replace with new Battery)  
5. Install new 9V battery into compartment and assure proper polarity of battery.  
(An alkaline type NEDA #1604 is recommended.)  
6. Install bottom case cover and secure with screws.  
Form number 61362  
rev 1 May 2002  
 
Page 5  
Replacing Fuse (refer to Figure 3)  
1. Disconnect the test leads and turn the range switch to “OFF”.  
2. Remove the three screws holding the bottom case cover  
3
Us a digital multimeter in low resistance {ohms} mode to check the two fuses  
mA input, 0.5A / 250V fast acting fuse.  
Amp input, 10A / 600V fast acting fuse  
4. Remove the defective fuse with the recommended fuse  
mA with : 0.5A/250V fast acting fuse Type LA-3895 is recommended.  
Amp with: 10A/600V fast acting fuse, Type LA-3897 is recommended  
5. Install bottom cover and secure with screws.  
Form number 61362  
rev 1 May 2002  
 

3M Musical Instrument Amplifier KD45 User Manual
Adder Technology Network Router CATxIP 1000 User Manual
ADT Security Services Home Security System BHS 3000C User Manual
Aico Smoke Alarm Ei169RF User Manual
Aiwa Speaker TS W60 User Manual
Alamo Lawn Mower 02968822P User Manual
Allied Telesis Network Router AR4155 User Manual
American Standard Outdoor Shower 2000500 User Manual
Beckett Weather Radio 7600WTM User Manual
Behringer Stereo System FEX800 User Manual